Catecholamines LC-MS/MS analysis kit

Catecholamines LC-MS/MS analysis kit

CE-IVD certified and suitable for any kind of LC-MS/MS instrument

 

The Zivak Catecholamines LC-MS/MS analysis kit can be used in combination with the Zivak-Multitasker as well as manually. Analysis run time for the manual version is just 16 minutes. Due to the efficient and quite smart procedure of the Zivak-Multitasker, the automated version guarantees also to receive results in 16 minutes per test, including sample preparation-, sample injection- and analysis run time. 

 

Quantitative LC-MS/MS analysis kit for Adrenalin, Noradrenalin and Dopamine in urine

Catecholamines are hormones produced by the adrenal glands, which are found on top of the kidneys. They are released into the blood during times of physical or emotional stress. The major catecholamines are dopamine, noradrenaline, and adrenaline. The measurements of catecholamines are important for determination of pheochromocytomas in symptomatic patients like patients with persistent hypertension. It is also used in order to help monitor for recurrence when a pheochromocytoma has been discovered and removed.

Zivak Catecholamines LC-MS/MS analysis kit was developed for accurate analysis of Adrenaline, Noradrenaline and Dopamine in urine samples which gives results in 16 minutes with minimal sample preparation. Main methods and procedures that have been selected are based on EN ISO 13485 and 98/79/EC.

Catecholamines LC-MS/MS analysis kit 0
LC-MS/MS Parameters

Device

LC-MS/MS Analysis System

Column

Zivak Catecholamines Urine LC-MS/MS Analytical column

Injection volume

10 µL

Gradient

 

00.00 min                            100 % A

08:30 min                            10   % A

11:00 min                            10   % A

11:01 min                            100 % A

16:00 min                            100 % A

Flow

0,35 mL/min

Ionization Mode:

ESI positive

CID Gas

2,4 mTorr

API Nebulizing Gas

50 psi

Scan Time

1,0 sec

SIM Width

1,5 amu

Drying Gas

35 psi

Drying Gas Temp.

350 °C

Needle

6000V

Shield

600V

Capillary

20V

Detector

1700 V

 

 

Analytical Performance

No

Analyte

LOD (µg/L)

LOQ (µg/L)

Accuracy (%)

Precision (%CV)

1

Adrenaline

0.4

1.2

97.5

3.2

2

Noradrenaline

0.6

1.8

98.5

2.6

3

Dopamine

0.3

0.9

98.3

2.4

 

Storage Conditions:

  • This analysis kit can be shipped at room temperature.
  • Standards and controls, Reagent 1, 2, 3 and 4 should be stored at 2-8 °C.
  • Standards and controls (after reconstitution) should be stored at - 20 °C.
  • All other components of the kit should be stored at between room temperature.
  • Reagent 2, 3 and 4 are volatile, please make sure that the lids of reagent bottles are closed tightly after use.
  • All components are guaranteed until expiry date when stored at recommended temperatures and used as described in these instructions.

 

Preparation of Calibrators:

  1. Add 8 ml of 0,2M HCl to lyophilised urine calibrator and dissolve.

  2. After reconstitution, Urine calibrator must be aliquoted into 120 µl volumes and stored at -20°C.

 

Preparation of Controls:

  1. Add 8 ml of 0,2M HCl to urine control and dissolve.
  2. After preparation, control must be aliquoted into 120 µl volumes and stored at -20 °C.

Sample Pretreatment (Manual):

  1. Take 100 mL of patient sample (or calibrators and controls) into the sample preparation tube.
  2. Add 100 mL of Reagent 1 and vortex for 10 seconds.
  3. Add 200 mL of Reagent 2 and vortex for 10 seconds.
  4. Add 200 mL of Reagent 3 and vortex for 10 seconds. Wait for 3 minutes and vortex again for 10 seconds. Wait for 1 minute.
  5. Add 200 mL of Reagent 4 and vortex for 10 seconds. After mixing wait for a minute.
  6. Centrifuge at 5.000 x rpm for 2 minutes.
  7. Take 200 mL the upper phase into the autosampler vial.
  8. Inject 10 mL to the LC-MS/MS system.

The prepared sample is stable at 2-8 °C for 24 hours.

 

Sample Pretreatent (Automated):

  1. Take 100 mL of patient sample (or calibrators and controls) into the Zivak Multitasker Sample Preparation vials
  2. Make sure that reagents and wash solution are connected to the reagent pump and purged properly. 
  3. Place the vials on the MULTITASKER system tray.

The prepared sample is stable at 2-8 °C for 24 hours.

 

Materials Supplied

ZV-3031-0200-20

Volume

Component

ZV-3031-02R1-20

1x 30ml

Reagent 1, Contains deuterated internal standard

ZV-3031-02R2-20

1x 50ml

Reagent 2, Contains base

ZV-3031-02R3-20

1x 50ml

Reagent 3, Contains base and derivatization reagent

ZV-3031-02R4-20

1x 50ml

Reagent 4, Contains organic extraction solvent

ZV-3031-02MA-20

2x 0,8 L

Mobile Phase A, Contains organic solvent

ZV-3031-02MB-20

1x 0,6 L

Mobile Phase B, Contains organic solvent

ZV-3031-02WB-20

1x 1 L

Washing Solution

ZV-3031-KK-20

1x 1pc

User Guide

 

Materials Required But Not Supplied

ZV-3031-0200-20

Volume

Component

ZV-3031-02S1-20

1x 8ml

Lyophilized  Urine Calibrator Level 1

ZV-3031-02K1-20

1x 8ml

Lyophilized  Urine Control Level 1

ZV-3031-02K2-20

1x 8ml

Lyophilized  Urine Control Level 2

ZV-3031-02C1-20

1 x 1 pcs

Zivak Catecholamines Urine LC-MS/MS Analytical column

ZS-2ML-8-1001

2x100pcs

2 ml glass autosampler vial

ZS-9004-0100-00

2x100pcs

2 ml glass autosampler vial caps pre-slit